On (b)(6) 2020, the patient was treated for an abdominal aortic aneurysm.The physician successfully implanted a gore® excluder® aaa endoprosthesis featuring c3® delivery system.He then implanted a gore® excluder® aaa endoprosthesis.After it was deployed, the physician went to remove the delivery system, and in doing so, the balloon valve on the gore® dryseal sheath with hydrophilic coating was ruptured.The delivery system, and sheath were removed, and a second gore® dryseal sheath with hydrophilic coating was used to complete the procedure.The patient tolerated the procedure.After the procedure, the sheath and delivery catheter were examined.The distal nose cone on the delivery system was found to have broken, with about 1 cm of the distal end separated from the rest of the nose cone.The sheath and delivery catheter will be returned for investigation.The physician is requesting a response.The physician reported that the neck of the aorta was heavily calcified.
|