MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL, 3-WAY SILICONE, TRICOT; STATLOCK DEVICE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 SL FOLEY SWIVEL, 3-WAY SILICONE, TRICOT; STATLOCK DEVICE Back to Search Results
Model Number FOL0105
Device Problem Disconnection (1171)
Patient Problems No Consequences Or Impact To Patient (2199); Insufficient Information (4580)
Event Date 09/30/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the 3 - way statlock came apart when the patient tried to use them and also stated that 3 out of 5 statlocks had the same issue.
 
Event Description
It was reported that the 3 - way statlock came apart when the patient tried to use them and also stated that 3 out of 5 statlocks had the same issue.
 
Manufacturer Narrative
The reported event was confirmed however the cause was unknown.Visual inspection noted one used foley statlock pad was received.Visual evaluation noted the clamp was disconnected and missing on return.This is a failure as the assembly must be completely bonded.Although the reported event was confirmed, a root cause could not be determined.A potential root cause for this failure could be glue selection incompatible with clamp base.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "statlock® foley stabilization device read carefully before use.Safety and efficacy considerations: single use.Do not alter the statlock® device or components.Procedure must be performed by trained personnel with knowledge of anatomical landmarks, safe technique and potential complications.Contents: package includes the statlock® device stabilization system and skin preparation pad.Indications for use: the statlock® device is a stabilization device for compatible catheters.Contraindications: known tape or adhesive allergies.Warnings and precautions: 1.Do not use the statlock® device where loss of adherence could occur, such as with a confused patient, diaphoretic or non-adherent skin, or when the access device is not monitored daily.2.Observe universal blood and body fluid precautions and infection control procedures, during application and removal of the statlock® device.3.Minimize catheter manipulation during application and removal of the statlock® device.4.Daily maintenance: a.The statlock® device should be assessed daily and changed when clinically indicated, at least every seven days.B.If pad becomes soiled, wash with soap/water, saline or hydrogen peroxide.Do not use alcohol or prepackaged bathing systems, which could lead to early lifting.C.If showering/bathing, cover with plastic wrap or waterproof dressing.D.Conduct skin assessment prior to application and repeat daily per facility protocol.E.Use clinical judgment on the removal of the statlock® stabilization device if the patient experiences any fluid shifts that may interfere with skin integrity.The statlock® stabilization device should be monitored daily and replaced when clinically indicated, at least every 7 days.The catheter insertion site should be treated per local policy recommendations.The statlock® device is contraindicated on patients with known tape and adhesive allergies.Alcohol and acetone may weaken the adhesive bond between the statlock® device pad and the skin.Care should be taken when using these solutions while performing catheter site care.Please consult product inserts and labels for any indications, contraindications, hazards, warnings, cautions and directions for use.For additional information on the statlock® foley stabilization device, please call your local bard representative.Instructions for use foley application technique prep 1.Place foley catheter into retainer.Directional arrow should point towards catheter tip, and balloon inflation arm should be next to the clamp hinge.2.Close lid, being careful to avoid pinching the catheter.3.Identify securement site by laying the device retainer on the front of the thigh, leaving 1 inch of catheter slack between insertion site and the statlock® device retainer.4.After placing the statlock® stabilization device off to the side, cleanse and degrease the securement site with alcohol per hospital policy.Let skin dry.5.Apply skin protectant, in direction of hair growth, to area larger than securement site.Allow to dry completely (10-15 seconds).6.Using permanent marker, write initials and date of application on the statlock® device anchor pad.Note: always secure catheter into the statlock® device retainer before applying adhesive pad on skin.Place and peel 7.Align the statlock® stabilization device over securement site leaving 1 inch of catheter slack.Make sure leg is fully extended.8.While holding the retainer to keep the pad in place, peel away paper backing, one side at a time and place tension-free on skin.Removal technique disengage 1.Open retainer by pressing release button with thumb, then lift to open.2.Remove foley catheter from the statlock® device.Dissolve 3.Wipe the edge of the pad using at least 5-6 alcohol pads until a corner lifts.Then continue to stroke undersurface of pad with alcohol to dissolve adhesive pad away from skin.Do not pull or force pad to remove." h11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SL FOLEY SWIVEL, 3-WAY SILICONE, TRICOT
Type of Device
STATLOCK DEVICE
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10734327
MDR Text Key214634039
Report Number1018233-2020-20433
Device Sequence Number1
Product Code EYJ
UDI-Device Identifier00801741076145
UDI-Public(01)00801741076145
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,distributor,other
Type of Report Initial,Followup
Report Date 12/29/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date01/28/2023
Device Model NumberFOL0105
Device Catalogue NumberFOL0105
Device Lot NumberJUEP0310
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/29/2020
Initial Date Manufacturer Received 09/30/2020
Initial Date FDA Received10/26/2020
Supplement Dates Manufacturer Received12/29/2020
Supplement Dates FDA Received01/19/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-