• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 17X119MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 17X119MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/05/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that debris was identified in the sterile packaging while investigating items in stock.No patients were involved.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.The reported event was confirmed.Evaluation of the returned product/photographs provided confirmed foreign debris is present inside the sterile packaging, and the sterile packaging remains sealed a complaint history review was conducted for part # (51-145170).The search identified 35 other complaint(s) for this device for the same or similar issue, within one (1) year prior to the notification date and thereafter.The search identified no other additional complaint(s) for the same lot (6695447).The likely condition of the device when it left zimmer biomet is non-conforming to specification.The root cause of the reported event is the operator not following the work instructions provided.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR MP T1 PPS 17X119MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10740070
MDR Text Key213187230
Report Number0001825034-2020-03904
Device Sequence Number1
Product Code LPH
UDI-Device Identifier00880304517363
UDI-Public(01)00880304517363
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,distri
Type of Report Initial,Followup
Report Date 01/05/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-145170
Device Lot Number6695447
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/27/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/05/2020
Initial Date FDA Received10/27/2020
Supplement Dates Manufacturer Received01/04/2021
Supplement Dates FDA Received01/05/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
-
-