Model Number 0570-0307
Device Problems Erratic or Intermittent Display (1182); No Display/Image (1183); Intermittent Loss of Power (4016)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/30/2020
Event Type  malfunction  
Manufacturer Narrative
The glidescope avl video baton 3-4 was returned to verathon for evaluation along with the corresponding glidescope avl 1.0 monitor.A verathon technical service representative evaluated both the returned video baton and monitor, but was unable to reproduce the reported issue.The glidescope avl 1.0 monitor and glidescope avl video baton 3-4 were returned to the customer on completion of the evaluation.No further investigation is required at this time.Verathon will continue to monitor for trends.
 
Event Description
The customer reported that during a patient procedure, using a glide scope avl video baton 3-4, it would intermittently reboot while the device was in use, and they would lose picture on the screen.When the biomed tested it out on his own he identified that the picture was jumpy when the baton was manipulated.No delay in the procedure, use of a backup device, or harm to the patient, or user was reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
GLIDESCOPE AVL VIDEO BATON 3-4
Type of Device
LARYNGOSCOPE, RIGID
Manufacturer (Section D)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA  V5C 5A9
Manufacturer (Section G)
VERATHON MEDICAL ULC
2227 douglas road
burnaby, british columbia V5C 5 A9
CA   V5C 5A9
Manufacturer Contact
corey kasbohm
20001 n creek pkwy
bothell, WA 98011-8218
4256295760
MDR Report Key10740649
MDR Text Key213207485
Report Number9615393-2020-00214
Device Sequence Number1
Product Code CCW
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Reporter Occupation Biomedical Engineer
Type of Report Initial
Report Date 10/01/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number0570-0307
Device Catalogue Number0570-0313
Device Lot NumberN/A
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/19/2020
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/01/2020
Initial Date FDA Received10/27/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured07/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-