• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: STRYKER MEDICAL-KALAMAZOO ISOLIBRIUM SUPPORT SURFACE - N; BED, PATIENT ROTATION, POWERED

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

STRYKER MEDICAL-KALAMAZOO ISOLIBRIUM SUPPORT SURFACE - N; BED, PATIENT ROTATION, POWERED Back to Search Results
Model Number 2971
Device Problem Device Slipped (1584)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Date 07/01/2020
Event Type  malfunction  
Manufacturer Narrative
This mdr is part of the fda voluntary malfunction summary reporting program.  the device was evaluated in the field and the issue was confirmed; there was a worn component.  the device was repaired and returned.  there was no remedial action taken.  this device is not labeled for single use.
 
Event Description
This report summarizes 1 malfunction event, where it was reported the device drifted.There was no patient involvement.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ISOLIBRIUM SUPPORT SURFACE - N
Type of Device
BED, PATIENT ROTATION, POWERED
Manufacturer (Section D)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer (Section G)
STRYKER MEDICAL-KALAMAZOO
3800 east centre avenue
portage MI 49002
Manufacturer Contact
kristen canter
3800 east centre avenue
portage, MI 49002
2693292100
MDR Report Key10741484
MDR Text Key213237948
Report Number0001831750-2020-01034
Device Sequence Number1
Product Code IKZ
UDI-Device Identifier07613327278392
UDI-Public07613327278392
Combination Product (y/n)N
Reporter Country CodeUS
Number of Events Reported1
Summary Report (Y/N)Y
Report Source Manufacturer
Source Type company representative
Reporter Occupation Other
Type of Report Initial
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2971
Device Catalogue Number2971000000
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/01/2020
Initial Date FDA Received10/27/2020
Was Device Evaluated by Manufacturer? Yes
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-