A2 date of birth: (b)(6) 1998.Device evaluated by mfr.: returned product consisted of the solent omni thrombectomy system.The pump assembly, effluent/supply line, shaft, tip, and spike line were visually inspected.Inspection presented no damage or irregularities to the pump assembly or any other part of the device.Functional testing was performed by placing the device in the angiojet ultra console.Testing consisted of running the complaint device through the full priming cycle and 120 seconds in thrombectomy mode.The device ran within normal range without any leaks from the pump assembly or anywhere else on the device and there were no errors/alarms from the console.
|