• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BECTON, DICKINSON & CO. (SPARKS) BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BECTON, DICKINSON & CO. (SPARKS) BD RAPID DETECTION OF SARS-COV-2 VERITOR; CORONAVIRUS ANTIGEN DETECTION SYSTEM Back to Search Results
Catalog Number 256082
Device Problem Incorrect, Inadequate or Imprecise Result or Readings (1535)
Patient Problem Test Result (2695)
Event Date 10/07/2020
Event Type  malfunction  
Manufacturer Narrative
Eua #: (b)(4).Medical device expiration date: unknown.A device evaluation is anticipated, but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.Device manufacture date: unknown.(b)(4).
 
Event Description
It was reported while testing for sars-cov-2 a false positive result was obtained.A repeat test was performed using a pcr test method and the result was negative.The customer stated the patient tested was asymptomatic.This test is not intended for use on asymptomatic patients and was therefore used off label.There was no report of patient impact.(b)(4).
 
Manufacturer Narrative
Eua#: (b)(4).Investigation summary: bd has received several customer complaints for false positive results, when using bd sars-cov-2 reagents for bd veritor¿ system.The current investigation concerns multiple lots of bd sars-cov-2 reagents for bd veritor¿ system.Bd takes a systematic approach to investigating false positive complaints that are received.This approach involves review of manufacturing batch history records, testing of retention samples, and testing of customer returned samples if applicable.Due to unknown lot# a dhr could not be performed.The investigation did not find a root cause for the false positive results that were observed.It is recommended that each customer review their workflow carefully to ensure that the package insert is being followed as written.The root cause is under investigation and will be documented in our quality system.Bd cannot confirm the complaint based on the investigation that was performed.A corrective and preventive action (capa#1878253) is already initiated to investigate the root cause and some mitigation actions are already being addressed.
 
Event Description
It was reported while testing for sars-cov-2 a false positive result was obtained.A repeat test was performed using a pcr test method and the result was negative.The customer stated the patient tested was asymptomatic.This test is not intended for use on asymptomatic patients and was therefore used off label.There was no report of patient impact.Eua (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD RAPID DETECTION OF SARS-COV-2 VERITOR
Type of Device
CORONAVIRUS ANTIGEN DETECTION SYSTEM
Manufacturer (Section D)
BECTON, DICKINSON & CO. (SPARKS)
7 loveton circle
sparks MD 21152
MDR Report Key10742312
MDR Text Key214759635
Report Number1119779-2020-00796
Device Sequence Number1
Product Code QKP
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type other,user facility
Type of Report Initial,Followup
Report Date 01/15/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Catalogue Number256082
Device Lot NumberUNKNOWN
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/07/2020
Initial Date FDA Received10/27/2020
Supplement Dates Manufacturer Received01/14/2021
Supplement Dates FDA Received01/15/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-