• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202599
Device Problems Break (1069); Material Separation (1562)
Patient Problems No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/09/2020
Event Type  Injury  
Event Description
It was reported that during a repair of medial collateral ligament of knee joint, the anchor was fractured.Backup device was available to complete the procedure.It is unknown if a delay happened.However, no patient complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10: h3,h6: the reported device was received for evaluation.There was a relationship found between the device and the reported event.A complaint history review found similar reported events.A review of device records showed there were no indications to suggest that the product did not meet manufacturing specification upon release for distribution.The instructions for use was reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A risk management review found that the reported failure was documented appropriately, and there were no indications to suggest the anticipated risk is not adequate.A review of the polymer, found that the storage requirements, material specifications, and applicable tests were appropriately specified.A material certificate of analysis was required for the raw material.A visual inspection of the returned device found that it is not in its original packaging.The device has been deployed, and the distal end of the anchor is fractured.There is debris on the anchor, sutures, and shaft of the device.Based on the condition of the product material found during visual inspection, additional material testing is not required.A review of the customer provided images found that the device has been deployed, and the anchor is in a pouch with the handheld device.The anchor is fractured.Although the tip of the anchor reportedly snapped, off in the patient during the rotator cuff procedure.Per communications, the tip was successfully removed.Based on the information provided, the procedure completed with a competitor¿s device.Since there was no delay or patient complications reported; no further clinical/medical assessment is warranted at this time.The complaint was confirmed, and the root cause was associated with unintended use of the device.Factors that could have contributed to the reported event include excessive force on the device, excessive torque on the device, attempted correction of a damaged device, off-axis insertion, improper preparation of the insertion site, or an inadvertent impact event inconsistent with normal use.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.
 
Event Description
It was reported that during a repair of medial collateral ligament of knee joint, the twinfix anchor was fractured.Pieces were successfully removed from the patient.Backup device was available to complete the procedure.Non-significant delay and no further complications were reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA PK 5.5MM W/2 UB-WHT & BL
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer (Section G)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
Manufacturer Contact
holly topping
7000 west william cannon drive
austin, TX 78735
5123913905
MDR Report Key10751964
MDR Text Key213638490
Report Number1219602-2020-01683
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010647474
UDI-Public03596010647474
Combination Product (y/n)N
Reporter Country CodeCH
PMA/PMN Number
K093228
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type Other,Foreign,User Facility,Company Representative
Reporter Occupation Other
Type of Report Initial,Followup
Report Date 12/06/2022
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date05/07/2024
Device Model Number72202599
Device Catalogue Number72202599
Device Lot Number2033256
Was Device Available for Evaluation? Device Returned to Manufacturer
Is the Reporter a Health Professional? No
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received10/28/2020
Supplement Dates Manufacturer Received12/05/2022
Supplement Dates FDA Received12/06/2022
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/07/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
Patient Age17 YR
Patient SexFemale
Patient Weight60 KG
-
-