Model Number TMICL12.6 |
Device Problems
No Apparent Adverse Event (3189); Ejection Problem (4009)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 08/10/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Unk, unk, unk, unk, unk.Work order search: no similar complaint type events were reported for units within the same lot.Claim# 431171.
|
|
Event Description
|
The reporter indicated that on (b)(6) 2020 a tmicl12.6, -8.50/1.0/089 (sphere/cylinder/axis), implantable collamer lens "was hung up in plunger.Lens not implanted." cause of the event is reported as "lens got hung up in the plunger." additional information has been requested but none has been forthcoming.If additional information is received a supplemental medwatch report will be submitted.
|
|
Manufacturer Narrative
|
H3 - device evaluation: lens was returned in a micro-centrifuge vial, in liquid.Visual inspection found no visible damage to the lens.Claim#: (b)(4).
|
|