MEDTRONIC SINGAPORE OPERATIONS ADAPTA; PULSE GENERATOR, PERMANENT, IMPLANTABLE Back to Search Results
Model Number ADDRL1
Device Problem Pocket Stimulation (1463)
Patient Problems Muscle Stimulation (1412); Device Overstimulation of Tissue (1991)
Event Date 09/30/2020
Event Type  Injury  
Manufacturer Narrative
If information is provided in the future, a supplemental report will be issued.
 
Event Description
It was reported that the patient felt twitches in their device pocket area.It was also reported that the pacing lead impedance had increased steadily since implant.The implantable pulse generator (ipg) and pacing lead remain in use and reprogramming will be considered during patient's next follow-up appointment with their physician.No further patient complications have been reported as a result of this event.
 
Event Description
It was further reported that mode switched occurred.Reprogramming was carried out and the ipg and pacing lead remain in use.
 
Manufacturer Narrative
Medtronic is submitting this report to comply with fda reporting regulations under 21 cfr parts 4 and 803.This report is based upon information obtained by medtronic, which the company may not have been able to fully investigate or verify prior to the date the report was required by the fda.Medtronic has made reasonable efforts to obtain more complete information and has provided as much relevant information as is available to the company as of the submission date of this report.This report does not constitute an admission or a conclusion by fda, medtronic, or its employees that the device, medtronic, or its employee caused or contributed to the event described in the report.In particular, this report does not constitute an admission by anyone that the product described in this report has any ¿defects¿ or has ¿malfunctioned¿.These words are included in the fda 3500a form and are fixed items for selection created by the fda to categorize the type of event solely for the purpose of regulatory reporting.Medtronic objects to the use of these words and others like them because of the lack of definition and the connotations implied by these terms.This statement should be included with any information or report disclosed to the public under the freedom of information act.Any required fields that are unpopulated are blank because the information is currently unknown or unavailable.A good faith effort will be made to obtain the applicable information relevant to the report.If information is provided in the future, a supplemental report will be issued.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ADAPTA
Type of Device
PULSE GENERATOR, PERMANENT, IMPLANTABLE
Manufacturer (Section D)
MEDTRONIC SINGAPORE OPERATIONS
49 changi south avenue 2
nasaco tech centre
singapore 48605 6
SG  486056
MDR Report Key10758926
MDR Text Key213864049
Report Number3008973940-2020-02716
Device Sequence Number1
Product Code NVZ
Combination Product (y/n)N
PMA/PMN Number
P980035
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,health professional
Type of Report Initial,Followup
Report Date 11/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Lay User/Patient
Device Expiration Date11/28/2017
Device Model NumberADDRL1
Device Catalogue NumberADDRL1
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/01/2020
Initial Date FDA Received10/29/2020
Supplement Dates Manufacturer Received11/26/2020
Supplement Dates FDA Received11/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
5038-58 LEAD; 5038-58 LEAD
Patient Outcome(s) Required Intervention;
-
-