Catalog Number 930700NS
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/02/2020
Event Type  malfunction  
Manufacturer Narrative
Pr 1879084 initial mdr.A follow up will be submitted if additional information becomes available.Patient problem code: 2692, device problem code: 2895.Na.
 
Event Description
It was reported that particulate was found.
 
Manufacturer Narrative
One photo was received by our quality team for evaluation.Upon visual inspection, it was observed that there were orange particles inside of the package; therefore, the incident could be verified.A review of the internal manufacturing device records and raw material history files for the reported lot numbers was performed and no recorded quality problems or rejections to this incident were found.Based on the investigation, the probable root cause could be from the manufacturing process.On 10.5ml applicators, particles may be created on rare occasions during the ultrasonic welding of the foam tip onto the applicator body which makes the foam brittle and pieces flake off.This failure mode is of foam origin not of foreign origin and it may be created as a byproduct of the ultrasonic welding process caused by the equipment.However, without a physical sample a full investigation could not be performed.This incident has been added to our database of reported incidents.Our business team regularly reviews the collected data for identification of emerging trends.
 
Event Description
It was reported that particulate was found.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
CHLORAPREP ONE-STEP
Type of Device
2% W/V CHLORHEXIDINE GLUCONATE/70% V/V ISOPROPYL ALCOHOL
Manufacturer (Section D)
CAREFUSION, INC
75 north fairway drive
vernon hills IL 60061
MDR Report Key10762738
MDR Text Key213881083
Report Number3004932373-2020-00116
Device Sequence Number1
Product Code KXF
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,other
Remedial Action Other
Type of Report Initial,Followup
Report Date 12/30/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date10/31/2022
Device Catalogue Number930700NS
Device Lot Number9313761
Initial Date Manufacturer Received 10/02/2020
Initial Date FDA Received10/30/2020
Supplement Dates Manufacturer Received10/02/2020
Supplement Dates FDA Received12/30/2020
Patient Sequence Number1
Patient Outcome(s) Other;
-
-