• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

BOSTON SCIENTIFIC CORPORATION VISIONIST X4 CRT-P; PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P) Back to Search Results
Model Number U228
Device Problems Signal Artifact/Noise (1036); Over-Sensing (1438); Connection Problem (2900); Material Integrity Problem (2978)
Patient Problems No Consequences Or Impact To Patient (2199); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/03/2020
Event Type  malfunction  
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was an attempted implant due to oversensing of electromagnetic interference (emi) noise that was suspected to be related to device header damage and set screw issues.There was no pacing inhibition.A different device same model was successfully implanted.No adverse patient effects were reported.The product has been received for analysis.This report will be updated upon completion of analysis.
 
Manufacturer Narrative
The returned device was thoroughly inspected and analyzed upon receipt at our post market quality assurance laboratory.Visual examination of the device header and case noted no anomalies.Pin gauge testing, designed to verify proper port dimensions, was completed.Each port measured as expected.The device was then exposed to simulated heart load conditions, and the pacing, and sensing functions were tested.Impedance testing was completed, and all measurements were within normal limits.The device operated appropriately with no interruptions in therapy output at the returned programmed settings.A series of electrical tests was also performed, and again, normal device function was observed.Analysis did not identify any device characteristics that would have caused or contributed to the reported clinical observations.
 
Event Description
It was reported that this cardiac resynchronization therapy pacemaker (crt-p) was an attempted implant due to oversensing of electromagnetic interference (emi) noise that was suspected to be related to device header damage and set screw issues.There was no pacing inhibition.A different device same model was successfully implanted.No adverse patient effects were reported.The product has been received for analysis.This report will be updated upon completion of analysis.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
VISIONIST X4 CRT-P
Type of Device
PULSE GENERATOR, PACEMAKER, IMPLANTABLE, WITH CARDIAC RESYNCHRONIZATION (CRT-P)
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
4100 hamline avenue north
saint paul MN 55112
MDR Report Key10765712
MDR Text Key214100311
Report Number2124215-2020-21700
Device Sequence Number1
Product Code NKE
UDI-Device Identifier00802526559471
UDI-Public00802526559471
Combination Product (y/n)N
PMA/PMN Number
P030005/S138
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 02/22/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date04/29/2022
Device Model NumberU228
Device Catalogue NumberU228
Device Lot Number752250
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer09/14/2020
Initial Date Manufacturer Received 09/03/2020
Initial Date FDA Received10/30/2020
Supplement Dates Manufacturer Received02/10/2021
Supplement Dates FDA Received02/23/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Age62 YR
Patient Weight91
-
-