ADAPTEC MEDICAL DEVICES LLC 3014271001 SENSICA UO MONITOR ICU; SENSICA UO MONITORICU Back to Search Results
Catalog Number SCCS1001-BARD
Device Problem Failure to Calibrate (2440)
Patient Problems No Patient Involvement (2645); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 10/07/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.The device was not returned.
 
Event Description
It was reported that the sensica unit was not calibrating.
 
Event Description
It was reported that the sensica unit was not calibrating.As per additional information received, biomed stated that after charging each of the systems, they were able to boot them with no errors.It was confirmed that the device was working properly.
 
Manufacturer Narrative
The reported issue was confirmed.The root cause of the reported issue could not be determined, as the reported issue could not be reproduced during evaluation.The device was not returned for evaluation.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "ifu for properly charging the device: when first using the sensica uo system, the internal back-up battery may require charging.Plug the system into a medical grade wall supply using the power cord provided, and allow a 20 hours to charge battery.To avoid battery drainage over time, is recommended to keep the system plugged into the wall during use whenever possible.The battery will recharge when the system is plugged into a wall supply.Cautions: during system start up and in general practice, plug the sensica uo system into a wall power supply whenever possible.After using the system on batter back-up, plug it back into the wall power supply recharging and to avoid system shut down due to a drained battery." h11: section a through f - the information provided by bard represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bard.H3 other text : the device was not returned.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SENSICA UO MONITOR ICU
Type of Device
SENSICA UO MONITORICU
Manufacturer (Section D)
ADAPTEC MEDICAL DEVICES LLC 3014271001
260 james jackson avenue
cary NC 27513
MDR Report Key10767542
MDR Text Key214108723
Report Number1018233-2020-20594
Device Sequence Number1
Product Code EXS
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 07/20/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Catalogue NumberSCCS1001-BARD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/07/2020
Initial Date FDA Received10/31/2020
Supplement Dates Manufacturer Received07/20/2021
Supplement Dates FDA Received08/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-