• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. ARTHROPIERCE INSTRUMENT CURVED LEFT; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. ARTHROPIERCE INSTRUMENT CURVED LEFT; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72201186
Device Problems Difficult to Open or Close (2921); Mechanical Jam (2983)
Patient Problems No Information (3190); No Clinical Signs, Symptoms or Conditions (4582)
Event Date 09/22/2020
Event Type  malfunction  
Event Description
It was reported that during arthroscopy, the arthropierce instrument's grabbing arm stopped closing in mid procedure.There was a delay of under 30 min, the procedure was completed with a change in technique using a competitor device (mitek cleverhook).No patient injury or other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review for the reported batch number concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.The instructions for use were reviewed and found to include conditions of off label use and technique specifics, as well as precautions and warnings related to the use of the device.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this investigation will be reopened for evaluation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ARTHROPIERCE INSTRUMENT CURVED LEFT
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10767698
MDR Text Key214093962
Report Number1219602-2020-01720
Device Sequence Number1
Product Code NBH
UDI-Device Identifier03596010605610
UDI-Public03596010605610
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 06/14/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72201186
Device Catalogue Number72201186
Device Lot Number50599989
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/09/2020
Initial Date FDA Received11/01/2020
Supplement Dates Manufacturer Received05/21/2021
Supplement Dates FDA Received06/14/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-