Envoy medical corp.(emc) was notified on 10/23/2020 of a patient whose sensor testing had no output immediately after a battery change procedure.A backed-out transducer lead was assumed, requiring a follow-up procedure.The follow-up procedure was completed immediately after the battery change procedure and a backed-out lead was confirmed.The backed-out lead was re-inserted to its proper position, and the patient's hearing was then found to be restored.The sp was not removed/replaced.No device deficiencies are alleged because the system was found to be functioning properly intraoperatively, and the issue is related to the lead backing-out.Patient/clinical history with emc: (b)(6).
|