• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY IRELAND - 9616671 ATTUNE CONV SHIM SZ7 10MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY IRELAND - 9616671 ATTUNE CONV SHIM SZ7 10MM; ATTUNE INSTRUMENTS : TIBIAL TRIALS Back to Search Results
Model Number 2545-00-940
Device Problem Component Missing (2306)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/19/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
During routine instrument checks yesterday, the mdrd technician found 4 items damaged in attune sets.Each of the impactor heads were cracked.Both the femoral and rp tibial impactor heads had vertical cracks on the back; while the fb tibial impactor had a vertical crack on the front.The other item that was damaged was a size 7/8 - 10mm shim; it was missing the metal rings for the 2 holes.The mdrd technician predicts that they went missing during a wash cycle - they were not found.None of the damage produce fragments or had a delay on surgery.All 4 items were taken out of the sets and are available for return.
 
Manufacturer Narrative
Product complaint # : (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.H10 additional narrative: added: d9.If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.Corrected: h3.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: examination of the returned device confirms the reported event.The need for corrective action is not indicated.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ATTUNE CONV SHIM SZ7 10MM
Type of Device
ATTUNE INSTRUMENTS : TIBIAL TRIALS
Manufacturer (Section D)
DEPUY IRELAND - 9616671
loughbeg ringaskiddy co.
cork
EI 
MDR Report Key10774518
MDR Text Key214373789
Report Number1818910-2020-23796
Device Sequence Number1
Product Code HWT
UDI-Device Identifier10603295135173
UDI-Public10603295135173
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup,Followup
Report Date 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number2545-00-940
Device Catalogue Number254500940
Device Lot NumberBFA1GGE
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/10/2020
Initial Date Manufacturer Received 10/20/2020
Initial Date FDA Received11/02/2020
Supplement Dates Manufacturer Received12/10/2020
01/12/2021
Supplement Dates FDA Received12/22/2020
01/13/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-