Model Number 71364009
Device Problem Break (1069)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/07/2020
Event Type  malfunction  
Event Description
It was reported that two head impactors were found cracked.There was no case involved.
 
Manufacturer Narrative
The associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device could not confirm the stated failure mode.The head impactor of the device has dents and nicks.The device was manufactured in 2020 and shows signs of moderate use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
FEMORAL HEAD IMPACTOR
Type of Device
PROSTHESIS, HIP, SEMI-CONSTRAIN, MET/CERAM/POLY, CEMENT OR NON-POROUS, UNCEMENT
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10774586
MDR Text Key214296959
Report Number1020279-2020-06029
Device Sequence Number1
Product Code LZO
UDI-Device Identifier03596010199256
UDI-Public03596010199256
Combination Product (y/n)N
PMA/PMN Number
K113789
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 12/14/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71364009
Device Catalogue Number71364009
Device Lot Number20CM25830
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/07/2020
Initial Date FDA Received11/02/2020
Supplement Dates Manufacturer Received11/20/2020
Supplement Dates FDA Received12/14/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-