• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

LIVANOVA DEUTSCHLAND S5 MAST ROLLER PUMP; CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 10-88-00
Device Problem Failure to Pump (1502)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/06/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the s5 mast roller pump.The incident occurred in (b)(6).The affected pump was requested back to the manufacturer site for investigation.The reported issue could not be reproduced.A runaway fault error message was identified by analysis of serial read-out.The processor board, power switch and the main cable will be replaced due to residues of dried fluids and corrosion.It cannot be ruled out that connection issue between parts may have led to the reported issue.
 
Event Description
Livanova (b)(4) received a report that a s5 mast roller pump could not be restarted at the end of the procedure.An error message was displayed on the system panel.After switching off/on the pump the error could be cleared and the pump could be restarted.The procedure could be completed without any further issue.There was no report of patient injury.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
S5 MAST ROLLER PUMP
Type of Device
CONSOLE, HEART-LUNG MACHINE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM  80309
Manufacturer (Section G)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich 80309
GM   80309
Manufacturer Contact
enrico greco
14401 w. 65th way
arvada, CO 80004
MDR Report Key10776552
MDR Text Key214314485
Report Number9611109-2020-00618
Device Sequence Number1
Product Code DWB
Combination Product (y/n)N
Reporter Country CodeNL
PMA/PMN Number
K062396
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Reporter Occupation Other Health Care Professional
Remedial Action Other
Type of Report Initial
Report Date 11/02/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number10-88-00
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/06/2020
Initial Date FDA Received11/03/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured10/20/2006
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-