• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 14X148MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC 133 T1 PPS SO 14X148MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problems Device Damaged Prior to Use (2284); Tear, Rip or Hole in Device Packaging (2385)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/13/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Product has been received by zimmer biomet and the investigation is in process.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported the 14 standard offset taperloc complete step was opened for usage and the inner sterilized packaging was compromised and the physician would not use it.It looked like bits of porous coating came off and inner plastic coating was scraped in numerous areas.Implant was not implanted.There was no impact to patient or delay in procedure.Attempts have been made and additional information on the reported event is unavailable at this time.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Complaint sample was evaluated and the reported event was confirmed.Evaluation of the returned product/photographs provided confirmed there is debris inside the sterile packaging which is consistent with the appearance of the porous coating inside the sterile barrier.Sterility has not been compromised.Review of the device history records identified no deviations or anomalies during manufacturing related to the reported event.The likely condition of the product when it left zimmer biomet was conforming to specification.The root cause of the reported event is likely to be due to transit damage.The device evaluation found no malfunction and the event did not contribute to injury, therefore this would not be considered a reportable event.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC 133 T1 PPS SO 14X148MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10777696
MDR Text Key214337972
Report Number0001825034-2020-03973
Device Sequence Number1
Product Code KWA
UDI-Device Identifier00880304491816
UDI-Public(01)00880304491816
Combination Product (y/n)N
PMA/PMN Number
K101086
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,health profession
Type of Report Initial,Followup
Report Date 12/22/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-103140
Device Lot Number6636170
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/23/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/13/2020
Initial Date FDA Received11/03/2020
Supplement Dates Manufacturer Received12/18/2020
Supplement Dates FDA Received12/22/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberNI
Patient Sequence Number1
-
-