Other, other text: additional information: a1, h6, h10: information was received on 12/17/2020 indicating that the failure did not happen in use with a patient, it occurred during routine maintenance.Device evaluation: one smiths medical cadd solis hpca pump was returned for analysis in a good condition.During analysis, all the 3 tests had a failed average accuracy delivery error which was out of the nominal allowed specifications.The customer?s concern regarding failed accuracy was unable to be confirmed.Delivery accuracy testing found the pump?s average delivery error to be within the published specification of +/-6%.Service recommended replacing the expulsor.Based on the evidence, the complaint was not confirmed, and the problem source of the reported event is unknown.
|