• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: DEPUY INTERNATIONAL LTD - 8010379 ASR UNI FEMORAL IMPL SIZE 53; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

DEPUY INTERNATIONAL LTD - 8010379 ASR UNI FEMORAL IMPL SIZE 53; ASR HIP SYSTEM : HIP METAL FEMORAL HEADS Back to Search Results
Catalog Number 999890253
Device Problems Naturally Worn (2988); Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Calcium Deposits/Calcification (1758); Cyst(s) (1800); Foreign Body Reaction (1868); Inflammation (1932); Pain (1994); Synovitis (2094); Osteolysis (2377); Test Result (2695); No Code Available (3191)
Event Date 10/17/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
It was reported that the trunnion broke off in head.Doi: unknown; dor: (b)(6) 2018; left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The reported event is considered one of the possible complications of joint replacement.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.For this investigation, no immediate action was required as no alleged deficiency with the device(s) was identified.- without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.- the device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.- from the event information received, it was not possible to determine the relationship of the device to the reported event.No product contribution was identified and no information received with this individual complaint indicates that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.
 
Event Description
After review of the medical records the patient was revised to address broken femoral stem, metallosis of the trochanteric bursa, synovitis, trochanteric bursitis and heterotopic bone.Operative note reported broken stem, metallosis wear and particles and larger pieces of metal present.There was a metallosis trochanteric bursa, neurolysis median nerve, osteolysis in the greater trochanter, in acetabulum there were 2 grey metal cysts in the bone.Doi: (b)(6) 2009, dor: (b)(6) 2018; left hip.
 
Manufacturer Narrative
Product complaint # (b)(4).This report is being submitted pursuant to the provisions of 21 cfr, part 803.This report may be based on information which has not been investigated or verified prior to the required reporting date.This report does not reflect a conclusion by depuy synthes joint reconstruction, or its employees that the report constitutes an admission that the product, depuy synthes joint reconstruction, or its employees caused or contributed to the potential event described in this report.Udi : (b)(4).If information is obtained that was not available for the initial report, a follow-up report will be filed as appropriate.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.No device associated with this report was received for examination.A worldwide lot specific complaint database search, or device history record (dhr) review, was not possible because the required lot code(s) was not provided.The reported event is considered one of the possible complications of joint replacement.The complaint information provided has been reviewed for complaint coding, medical device reporting, and other data required by the complaint system.Follow-up for additional event information, if applicable, was conducted utilizing work instruction wi-7915 appendix a.For this investigation, no immediate action was required as no alleged deficiency with the device(s) was identified.Without the physical complaint sample(s) associated with this report, it was not possible to determine if the device(s) failed to meet specification(s) at the time it was released for distribution.The device(s) associated with this event were used in the treatment of the patient as prescribed by the presiding surgeon.From the event information received, it was not possible to determine the relationship of the device to the reported event.No product contribution was identified and no information received with this individual complaint indicates that a broader investigation or corrective action was necessary.Depuy synthes considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation may be re-opened as necessary.Device history lot : previous investigations that have included manufacturing record evaluations (mre) since the asr platform was launched have shown no indication of deviations or anomalies with regard to material, manufacturing or inspection.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ASR UNI FEMORAL IMPL SIZE 53
Type of Device
ASR HIP SYSTEM : HIP METAL FEMORAL HEADS
Manufacturer (Section D)
DEPUY INTERNATIONAL LTD - 8010379
st. anthony's road
leeds LS11 8DT
UK  LS11 8DT
MDR Report Key10782347
MDR Text Key214518070
Report Number1818910-2020-23881
Device Sequence Number1
Product Code KWA
Combination Product (y/n)N
PMA/PMN Number
K040627
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,consum
Remedial Action Recall
Type of Report Initial,Followup,Followup,Followup
Report Date 10/20/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date03/31/2019
Device Catalogue Number999890253
Device Lot Number2868843
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/20/2020
Initial Date FDA Received11/03/2020
Supplement Dates Manufacturer Received11/06/2020
03/16/2021
08/05/2021
Supplement Dates FDA Received11/08/2020
03/31/2021
08/10/2021
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberZ-1749/1816-2011
Patient Sequence Number1
Treatment
ADAPTER SLEEVES 12/14 +8; ASR ACETABULAR CUPS 60; PRODIGY X-LG STAT LT 16.5MM; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL AUGMENT; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL STEM; UNKNOWN HIP ACETABULAR CUP; UNKNOWN HIP FEMORAL AUGMENT; UNKNOWN HIP FEMORAL HEAD; UNKNOWN HIP FEMORAL STEM
Patient Outcome(s) Required Intervention;
Patient Age66 YR
-
-