SMITH & NEPHEW, INC. TWINFIX ULTRA TI 5.5 W/3 UB; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
|
Back to Search Results |
|
Model Number 72202897 |
Device Problem
Detachment of Device or Device Component (2907)
|
Patient Problem
No Information (3190)
|
Event Date 10/09/2020 |
Event Type
malfunction
|
Event Description
|
It was reported that during the surgery when the anchor was being screwed, the stem of the handle was stripped in the distal part.This caused a disconnection of the anchor and it was removed from the patient's anatomy.No additional bone hole was required.No patient injuries or delay reported.Smith and nephew back-up device was available to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
|
Manufacturer Narrative
|
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found:- incomplete anchor insertion may result in poor anchor performance - use of excessive force during insertion can cause failure of the suture anchor or insertion device.A two-finger ao technique should be used to insert the anchor.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
|
|
Search Alerts/Recalls
|
|
|