SMITH & NEPHEW, INC. TWINFIX ULTRA TI 5.5 W/3 UB; FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE Back to Search Results
Model Number 72202897
Device Problem Detachment of Device or Device Component (2907)
Patient Problem No Information (3190)
Event Date 10/09/2020
Event Type  malfunction  
Event Description
It was reported that during the surgery when the anchor was being screwed, the stem of the handle was stripped in the distal part.This caused a disconnection of the anchor and it was removed from the patient's anatomy.No additional bone hole was required.No patient injuries or delay reported.Smith and nephew back-up device was available to complete the surgery.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
The reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was an isolated event.A review of risk management files found that the reported failure was documented appropriately.A review of the instructions for use found:- incomplete anchor insertion may result in poor anchor performance - use of excessive force during insertion can cause failure of the suture anchor or insertion device.A two-finger ao technique should be used to insert the anchor.A relationship, if any, between the subject device and the reported event could not be determined.Please refer to the instructions for use for recommendations on proper use of the device and potential troubleshooting methods to prevent future reoccurrence of the reported event.No containment or corrective actions are recommended at this time.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TWINFIX ULTRA TI 5.5 W/3 UB
Type of Device
FASTENER, FIXATION, NONDEGRADABLE, SOFT TISSUE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
130 forbes blvd.
mansfield MA 02048
MDR Report Key10785129
MDR Text Key214552117
Report Number1219602-2020-01765
Device Sequence Number1
Product Code MBI
UDI-Device Identifier03596010652317
UDI-Public03596010652317
Combination Product (y/n)N
PMA/PMN Number
K100159
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/08/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72202897
Device Catalogue Number72202897
Device Lot Number2049856
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received11/04/2020
Supplement Dates Manufacturer Received12/04/2020
Supplement Dates FDA Received12/08/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-