Model Number N/A |
Device Problem
Material Deformation (2976)
|
Patient Problems
No Consequences Or Impact To Patient (2199); No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/16/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
Zimmer biomet (b)(4).The device will not be returned for analysis as it was discarded; however, an investigation of the reported event is in progress.Once the investigation is completed, a supplemental medwatch 3500a will be submitted.
|
|
Event Description
|
It was reported the sternal screw stripped easily upon insertion during a primary sternotomy.When the screw was being tightened with a manual driver, the screw stripped easily and deformed the screw head.The screw head was removed using a chisel and mallet while the screw body was removed using pick-ups.The screw was replaced with a different screw one size smaller.The malfunction resulted in a five (5) minute surgical delay.No adverse events have been reported as a result of the malfunction.
|
|
Manufacturer Narrative
|
This follow-up report is being submitted to relay additional information.No product was returned or pictures provided; visual and dimensional evaluations could not be performed.Lot identification is necessary for review of device history records, lot identification was not provided.A definitive root cause cannot be determined.If any further information is found which would change or alter any conclusions or information, a supplemental report will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|