ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG LEAD Back to Search Results
Model Number MN20450-50A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Pain (1994)
Event Date 10/14/2020
Event Type  Injury  
Manufacturer Narrative
Date of event is estimated.Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined.
 
Event Description
It was reported the patient was experiencing pain at the lead site (at t12 vertebral level and upper lateral thigh).X-rays were taken which indicated there may have been some extra loops in the target foramen causing irritation.In turn, surgical intervention was undertaken on (b)(6) 2020 wherein physician decided to pull some slack back that may be sitting in the foramen of t12 and l1.This addressed the issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer (Section G)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
Manufacturer Contact
andrea deitz
6901 preston road
plano, TX 75024
9723098000
MDR Report Key10786425
MDR Text Key214592308
Report Number1627487-2020-33283
Device Sequence Number1
Product Code PMP
Combination Product (y/n)N
Reporter Country CodeAS
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date10/27/2021
Device Model NumberMN20450-50A
Device Lot Number7234780
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/14/2020
Initial Date FDA Received11/04/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured10/28/2019
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Other;
-
-