Model Number BI70000028120 |
Device Problem
Device Sensing Problem (2917)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/28/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
A medtronic representative went to the site to test the equipment.Testing revealed that the issue could not be replicated.They adjusted door sensor as a precaution and verified doors are closing properly and the pendant shows doors closed.If information is provided in the future, a supplemental report will be issued.
|
|
Event Description
|
Medtronic received information regarding an imaging system that was being used for a procedure.It was reported that the system would not take a spin and gave no yellow or green lights.The pendant stated the door was still open, although it was closed.The rt banged on the side of the gantry and it engaged the door sensor and they were able to proceed.There was a reported delay of less than an hour and no impact to patient outcome.
|
|
Search Alerts/Recalls
|