Model Number M00546610
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Perforation (2001)
Event Date 10/01/2020
Event Type  Injury  
Manufacturer Narrative
(date of event): the exact date of the event is unknown.The provided event date (b)(6) 2020 was chosen as a best estimate based on the date that the manufacturer became aware of the event.The complainant was unable to report the lot number; therefore the manufacture date and expiration date are unknown.(b)(4).The complainant indicated that the device was disposed and will not be returned for evaluation; therefore, a failure analysis of the complaint device could not be completed.If any further relevant information is identified, a supplemental medwatch will be filed.
 
Event Description
It was reported to boston scientific corporation that a spyscope ds ii was used in the bile duct during an endoscopic sphincterotomy (est) procedure performed on an unknown date.Reportedly, there were no issues noted with the spyscope ds ii during the procedure.According to the complainant, during the procedure, the area near the papillae was checked and a large perforation was observed in the nipple est area.Reportedly, the physician thinks that the papilla vater was small and the est cut was also small, so the perforation was probably caused as the spyscope ds ii was inserted into and removed from the papilla vater many times.Upon discovering the perforation, a clip was used immediately, endoscopic nasobiliary drainage (enbd) was placed, and an nasogastric tube was also placed.The patient's condition at the conclusion of the procedure was reported to be recovered.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SPYSCOPE DS II ACCESS & DELIVERY CATHETER
Type of Device
CHOLEDOCHOSCOPE AND ACCESSORIES, FLEXIBLE/RIGID
Manufacturer (Section D)
BOSTON SCIENTIFIC CORPORATION
300 boston scientific way
marlborough MA 01752
Manufacturer (Section G)
BOSTON SCIENTIFIC CORPORATION
780 brookside drive
spencer IN 47460
Manufacturer Contact
carole morley
300 boston scientific way
marlborough, MA 01752
5086834015
MDR Report Key10791121
MDR Text Key214751610
Report Number3005099803-2020-05116
Device Sequence Number1
Product Code FBN
UDI-Device Identifier08714729965404
UDI-Public08714729965404
Combination Product (y/n)N
Reporter Country CodeJA
PMA/PMN Number
K142922
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Physician
Type of Report Initial
Report Date 11/04/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model NumberM00546610
Device Catalogue Number4661
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/14/2020
Initial Date FDA Received11/05/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-