• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. URETERO-RENO FIBERSCOPE Back to Search Results
Model Number URF-P6R
Device Problems Leak/Splash (1354); Component Missing (2306)
Patient Problem No Patient Involvement (2645)
Event Type  malfunction  
Manufacturer Narrative
Device was received and evaluated.Evaluation determined that the device was found with a leak on the ¿a-rubber¿ unit.The distal end plastic c-body unit was observed with a leak and multiple broken fibers were found on the image.In addition, the angulation unit was found not working.The identified parts needs to be replaced.The unit was placed for repair.The reported failure was confirmed.Root cause likely attributed to users maintenance and or handling issue.
 
Event Description
It was reported that during reprocessing, the device was observed with a leak and there was missing part on the lens.There was no patient involvement on this report.No user injury reported.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
URETERO-RENO FIBERSCOPE
Type of Device
URETERO-RENO FIBERSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10796070
MDR Text Key225182985
Report Number8010047-2020-08585
Device Sequence Number1
Product Code FGB
UDI-Device Identifier04953170340833
UDI-Public04953170340833
Combination Product (y/n)N
Reporter Country CodeCA
PMA/PMN Number
K172298
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Reporter Occupation Other
Type of Report Initial
Report Date 11/05/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberURF-P6R
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer08/31/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 08/31/2020
Initial Date FDA Received11/05/2020
Was Device Evaluated by Manufacturer? Yes
Date Device Manufactured06/06/2018
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-