The reported events ocd (over current detection) and sosd (shorted output state detection) were both confirmed.The analysis indicated that there was damage to the high voltage output circuitry of the device.The cause of the damage was high voltage delivery into a low impedance load.This low impedance is consistent with a lead-to-can arc and, therefore, is not considered a device malfunction.The reported event of reset was confirmed by analysis.The cause of the reset was found to be due to a por (power on reset).The cause of the por is consistent with external defibrillation and, therefore, is also not considered a device malfunction.Interrogation of the device revealed that the device was above elective replacement indicator when received.All low voltage device functions were normal.
|