• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA GASTROINTESTINAL VIDEO SCOPE; GASTROINTESTINAL VIDEOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

OLYMPUS MEDICAL SYSTEMS CORP. EVIS LUCERA GASTROINTESTINAL VIDEO SCOPE; GASTROINTESTINAL VIDEOSCOPE Back to Search Results
Model Number GIF-FQ260Z
Device Problem Mechanical Problem (1384)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/09/2020
Event Type  malfunction  
Manufacturer Narrative
Omsc reviewed the manufacturing history (dhr) of the device and confirmed no irregularity.The repair history and the repair history shows that the angulation wire the extra screw were assembled in the repair in may 2019.Therefore, omsc guessed that the reported event was caused by olympus repair process.If additional information becomes available, this report will be supplemented.
 
Event Description
Olympus found an extra screw and a piece of the angulating wire in the control body of the device during the repair process.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVIS LUCERA GASTROINTESTINAL VIDEO SCOPE
Type of Device
GASTROINTESTINAL VIDEOSCOPE
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10799167
MDR Text Key224630304
Report Number8010047-2020-08612
Device Sequence Number1
Product Code FDS
Combination Product (y/n)N
Reporter Country CodeUK
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type distributor,user facility
Reporter Occupation Non-Healthcare Professional
Type of Report Initial
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator No Information
Device Model NumberGIF-FQ260Z
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/09/2020
Initial Date FDA Received11/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured05/27/2009
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Unknown
Patient Sequence Number1
-
-