Model Number OER-4
Device Problem Contamination /Decontamination Problem (2895)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/09/2020
Event Type  malfunction  
Manufacturer Narrative
The device has not been returned to omsc for evaluation.Omsc reviewed the manufacturing history (dhr) of the device and confirmed no irregularity.The instruction manual of the device recommends that the water filter be replaced once a month.Omsc will inform user facility about the contents of the instruction manual and replacement of consumables.If additional information becomes available, this report will be supplemented.
 
Event Description
The user facility reported that they have not replaced the water filter of the device for two years.Therefore, there is a possibility that this device did not reprocess endoscopes sufficiently.There was no report of patient injury associated with this event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
OER-4 100V
Type of Device
ENDOSCOPE REPROCESSOR
Manufacturer (Section D)
OLYMPUS MEDICAL SYSTEMS CORP.
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8 507
JA  192-8507
Manufacturer Contact
kazutaka matsumoto
2951 ishikawa-cho
hachioji-shi, tokyo-to 192-8-507
JA   192-8507
426425177
MDR Report Key10799250
MDR Text Key224629899
Report Number8010047-2020-08615
Device Sequence Number1
Product Code FEB
Combination Product (y/n)N
Reporter Country CodeJA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type user facility
Reporter Occupation Other
Type of Report Initial
Report Date 11/06/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberOER-4
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/09/2020
Initial Date FDA Received11/06/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured12/02/2016
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
-
-