H3.Device evaluation by manufacturer: the aquabeam handpiece and associated aquabeam scope were returned for investigation.Visual inspection confirmed that the scope and handpiece were stuck together; the handpiece was bent at the probe.The probe was bent at approximately 6.26in from the manifold.Cracks were observed on the manifold adhesive of the handpiece.A review of the aquabeam robotic system's log file was reviewed, which confirmed a delay of approximately 10 minutes and 23 seconds when the handpiece and scope were replaced.A review of the device history records (dhr) was conducted for aquabeam robotic system/serial number (b)(6) and aquabeam handpiece/lot number 20c00117, which confirmed that there were no nonconformances, failures, discrepancies, or missed steps during the manufacturing process that could be related to the reported event.The review indicated that the system/handpiece met all design and manufacturing specifications when released for distribution.A review for similar complaints confirmed no other similar events reported to procept biorobotics.A revew of the aquabeam robotic system user manual, um0101 rev f, states: 11.2.5 sterile: aquabeam handpiece and aquabeam scope setup states: -hold the distal end of the scope tube tip approximately 1 inch (2.54cm) from the fully proximal position and continue advancing the aquabeam scope forward until it is properly engaged with the aquabeam handpiece and then rotate the proximal key alignment adapter so that the dimple on the proximal key alignment adapter is facing up.-an audible click should be heard when the aquabeam scope is securely engaged with the aquabeam handpiece.Note: do not use excessive force to advance the aquabeam scope.If resistance is felt, gently rotate the aquabeam scope clockwise and counterclockwise while simultaneously applying forward pressure.A revew of the aquabeam robotic system instructions for use, ifu0101-00 rev e, states: 5.1.Precautions: general: if excessive resistance is encountered during aquabeam scope positioning, reposition the aquabeam handpiece to minimize tenting the prostate in order to reduce the chance of damaging the aquabeam scope.The root cause of the reported event was determined to be user-related.There were cracks in the manifold adhesive from a compressive load due to tenting the prostate during the procedure.Mishandling of the scope during setup and positioning can increase the chances of failures.The aquabeam robotic system's user manual and instructions for use contains adequate to address this issue.Submission of this report does not constitute an admission that the manufacturer's product caused or contributed to the event.
|