• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SMITH & NEPHEW, INC. SVCE REPL LIGHT SOURCE 500XL XENON; IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE Back to Search Results
Model Number 72200568S
Device Problems Protective Measures Problem (3015); Unexpected Shutdown (4019)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/13/2020
Event Type  malfunction  
Event Description
It was reported that during an arthroscopic procedure, the light source suddenly shut off and presented e1, e3 and e13 error.The procedure was successfully completed with the same device with no surgical delay, the error disappeared when rebooting the device.No patient injuries and complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3,h6: the reported device, used in treatment, was received for evaluation.There was a relationship found between the returned device and the reported incident.The complaint of errors and power loss was confirmed.The ballast fan shuts down when the power button is turned on.Then the ballast overheats.Cause of errors and power loss is a defective electronic component.The complaint investigation has concluded the cause of the failure to be a defective electronic component.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review concluded this was a repeat issue.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
SVCE REPL LIGHT SOURCE 500XL XENON
Type of Device
IMAGE, ILLUMINATION, FIBEROPTIC, FOR ENDOSCOPE
Manufacturer (Section D)
SMITH & NEPHEW, INC.
76 s. meridian ave.
oklahoma city OK 73107
MDR Report Key10806974
MDR Text Key215279786
Report Number1643264-2020-01718
Device Sequence Number1
Product Code FFS
UDI-Device Identifier03596010650504
UDI-Public03596010650504
Combination Product (y/n)N
PMA/PMN Number
K994084
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number72200568S
Device Catalogue Number72200568S
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer12/16/2020
Initial Date Manufacturer Received 10/14/2020
Initial Date FDA Received11/08/2020
Supplement Dates Manufacturer Received12/22/2020
Supplement Dates FDA Received12/26/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-