• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ARTHROCARE CORP. QFIX 1.8 MINI SUTURE ANCHOR DISP KIT XL; ACCESSORIES,ARTHROSCOPIC

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ARTHROCARE CORP. QFIX 1.8 MINI SUTURE ANCHOR DISP KIT XL; ACCESSORIES,ARTHROSCOPIC Back to Search Results
Model Number 72290126
Device Problem Break (1069)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2020
Event Type  malfunction  
Event Description
It was reported that during shoulder surgery, the first qfix mini suture anchor had no implant on the end and the second had a broken drill.Nothing fell inside the patient.A backup device was available to complete the procedure with no delay or other complications.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H10 h3, h6: the reported device, used in treatment, was not returned to the designated complaint unit for independent evaluation, thus visual inspection and functional testing could not be performed.A review of the device history records showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.A complaint history review found no related failures; this failure mode will be trended to assess for any necessary corrective actions.A relationship, if any, between the subject device and the reported event could not be determined.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.If the product associated with this event is returned at a future date, this evaluation will be reopened for investigation.Internal complaint reference: (b)(4).
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
QFIX 1.8 MINI SUTURE ANCHOR DISP KIT XL
Type of Device
ACCESSORIES,ARTHROSCOPIC
Manufacturer (Section D)
ARTHROCARE CORP.
7000 w. william cannon
austin TX 78735
MDR Report Key10807362
MDR Text Key215274827
Report Number3006524618-2020-00966
Device Sequence Number1
Product Code NBH
UDI-Device Identifier00885556662649
UDI-Public00885556662649
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,user faci
Type of Report Initial,Followup
Report Date 12/11/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/07/2023
Device Model Number72290126
Device Catalogue Number72290126
Device Lot Number2046649
Was Device Available for Evaluation? Yes
Initial Date Manufacturer Received 10/16/2020
Initial Date FDA Received11/09/2020
Supplement Dates Manufacturer Received12/09/2020
Supplement Dates FDA Received12/11/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-