Catalog Number 00119301200 |
Device Problem
Fracture (1260)
|
Patient Problem
No Information (3190)
|
Event Date 10/07/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
(b)(4).Report source: (b)(6).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up report will be submitted.Product not returned.
|
|
Event Description
|
It was reported that the instrument fractured.
|
|
Manufacturer Narrative
|
Visual examination of the returned product identified the end that engages the lag screw is cracked.Material hardness is conforming to print specification and the device has a potential field age of approximately 16 years 6 months.Device history record was reviewed and no discrepancies were found.Root cause was unable to be determined.If any further information is found which would change or alter any conclusions or information, a supplemental will be filed accordingly.Zimmer biomet will continue to monitor for trends.
|
|
Event Description
|
No further event information available at the time of this report.
|
|