MAUDE Adverse Event Report: ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 16X117MM; PROSTHESIS, HIP

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ZIMMER BIOMET, INC. TPRLC XR MP T1 PPS 16X117MM; PROSTHESIS, HIP Back to Search Results
Model Number N/A
Device Problem Device Contaminated During Manufacture or Shipping (2969)
Patient Problem No Patient Involvement (2645)
Event Date 10/20/2020
Event Type  malfunction  
Manufacturer Narrative
(b)(4).Report source: (b)(4).Customer has indicated that the product is in process of being returned to zimmer biomet for investigation.Once the investigation has been completed, a follow-up mdr will be submitted.
 
Event Description
It was reported that during investigation in (b)(6) warehouse, circulated items (our stock) there was debris in sterile package.No patient involvement.No additional information.
 
Manufacturer Narrative
This follow-up report is being submitted to relay additional information.Visual examination of the returned product/provided pictures identified strike marks on the strike plate and light damages such as nicks and scratches.Inspection confirms the impactor pad has fractured and all the pieces were returned.Part and lot verified.Product sent to sem for further analysis: the fracture likely initiated at the thread interface.Fracture surface artifacts of hackle marks and river lines, coupled with the approximately 45 degree fracture plane, suggest torsional overload fracture, possibly associated with overtightening.Review of the device history record(s) identified no deviations or anomalies during manufacturing related to the reported event.Reported event is not related to a combination of products; therefore a compatibility review is not applicable.Review of complaint history identified additional similar complaints for the reported item(s) and no additional complaints for the reported part and lot combination(s).Complaints are monitored through monthly complaint review (reference sop040001 and wi040015) in order to identify potential adverse trends.Medical records were not provided.A definitive root cause cannot be determined.No corrective actions, preventive actions, or field actions resulted after investigation of this event.
 
Event Description
No further event information available at the time of this report.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
TPRLC XR MP T1 PPS 16X117MM
Type of Device
PROSTHESIS, HIP
Manufacturer (Section D)
ZIMMER BIOMET, INC.
56 e. bell drive
warsaw IN 46582
MDR Report Key10812065
MDR Text Key215381467
Report Number0001825034-2020-04039
Device Sequence Number1
Product Code LPH
Combination Product (y/n)N
PMA/PMN Number
K120030
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,foreig
Type of Report Initial,Followup
Report Date 12/28/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberN/A
Device Catalogue Number51-145160
Device Lot Number6490248
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/06/2020
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/20/2020
Initial Date FDA Received11/09/2020
Supplement Dates Manufacturer Received12/23/2020
Supplement Dates FDA Received12/30/2020
Is This a Reprocessed and Reused Single-Use Device? No
Removal/Correction NumberN/A
Patient Sequence Number1
-
-