Model Number 801188 |
Device Problems
Calibration Problem (2890); Output Problem (3005)
|
Patient Problem
No Known Impact Or Consequence To Patient (2692)
|
Event Date 10/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
The field service representative (fsr) duplicated the calibration failure.He replaced the oxygen (o2) sensor.The unit operated to the manufacturer's specifications.The suspect device was sent back to the manufacturer for further evaluation.
|
|
Event Description
|
It was reported that the electronic patient gas system (epgs) displayed a "failure to calibrate gas" error message.No other details regarding the nature of this event were provided.
|
|
Manufacturer Narrative
|
During laboratory analysis, the product surveillance technician could not duplicate the reported complaint.The returned oxygen sensor was installed into a lab use only electronic patient gas system and calibrated three consecutive times.
|
|
Manufacturer Narrative
|
The reported complaint was confirmed.Per the data log review on (b)(6) 2020, each time the gas system calibration was attempted it failed with a 'o2 sensor out of range at calibration' (sensor value = 0) message.This indicated a bad o2 sensor or a connection issue to the sensor.The data log confirms the complaint.If additional information becomes available on this complaint that would alter the facts and/or conclusion, a supplemental report will be filed accordingly.
|
|