Model Number V60 |
Device Problems
Defective Alarm (1014); Operating System Becomes Nonfunctional (2996)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Type
malfunction
|
Manufacturer Narrative
|
Date of event: (b)(6) 2020.Date of report: 09nov2020.
|
|
Event Description
|
It was reported to philips that the device had primary alarm failed alarm.The screen was frozen while a patient was on the unit and the alarm didn't stop sounding.The device was reported as being in use at the time of the event on a patient.The initial reporter confirmed that there was no adverse patient impact.
|
|
Manufacturer Narrative
|
G4:01mar2021.B4:03mar2021.H11:g5:k102985.H10: the customer had a standby ventilator which was immediately connected to patient and there was no delay in therapy.A philips service engineer (se) was dispatched to the customer site.The se confirmed the reported issue was caused by a 35v supply and auxiliary alarm supply failure.The internal power of 35v had decreased to around 4v.The se replaced the motor control printed circuit board assembly (mc pcba) to resolve the issue and bring the device back to functionality.Following the repair, the device was placed back into service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|
Manufacturer Narrative
|
G4:02mar2021.B4:16mar2021.There was no delay in therapy and no medical intervention was required.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
|
|