Model Number V60
Device Problems Defective Alarm (1014); Operating System Becomes Nonfunctional (2996)
Patient Problem No Clinical Signs, Symptoms or Conditions (4582)
Event Type  malfunction  
Manufacturer Narrative
Date of event: (b)(6) 2020.Date of report: 09nov2020.
 
Event Description
It was reported to philips that the device had primary alarm failed alarm.The screen was frozen while a patient was on the unit and the alarm didn't stop sounding.The device was reported as being in use at the time of the event on a patient.The initial reporter confirmed that there was no adverse patient impact.
 
Manufacturer Narrative
G4:01mar2021.B4:03mar2021.H11:g5:k102985.H10: the customer had a standby ventilator which was immediately connected to patient and there was no delay in therapy.A philips service engineer (se) was dispatched to the customer site.The se confirmed the reported issue was caused by a 35v supply and auxiliary alarm supply failure.The internal power of 35v had decreased to around 4v.The se replaced the motor control printed circuit board assembly (mc pcba) to resolve the issue and bring the device back to functionality.Following the repair, the device was placed back into service.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Manufacturer Narrative
G4:02mar2021.B4:16mar2021.There was no delay in therapy and no medical intervention was required.Submission of a report does not constitute an admission that medical personnel, user facility, importer, distributor, manufacturer, or product caused or contributed to the event.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
V60 VENTILATOR
Type of Device
VENTILATOR, CONTINUOUS, MINIMAL VENTILATORY SUPPORT,FACILITY USE
MDR Report Key10814132
MDR Text Key217151552
Report Number2031642-2020-04142
Device Sequence Number1
Product Code MNT
Combination Product (y/n)N
PMA/PMN Number
K082660
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,user facility
Type of Report Initial,Followup,Followup
Report Date 10/26/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model NumberV60
Was Device Available for Evaluation? Yes
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/26/2020
Initial Date FDA Received11/09/2020
Supplement Dates Manufacturer Received10/26/2020
10/26/2020
Supplement Dates FDA Received03/03/2021
03/16/2021
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured04/22/2019
Is the Device Single Use? No
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Reuse
Patient Sequence Number1
Treatment
UNKNOWN PATIENT CIRCUIT, MASK, AND HUMIDIFIER.
-
-