It was reported that during the procedure, the first pass mini upper jaw broke and fell inside the patient.The broken piece was removed.The procedure was successfully completed without delay using a back-up device.No other complications were reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
|
H10 internal complaint reference: (b)(4).H3, h6 the reported device, used in treatment, was returned for evaluation.There was a relationship found between the reported incident and the returned device.A review of the device history records for the reported lot number showed there were no indications to suggest that the product did not meet manufacturing specification or would not be able to perform as intended.Visual evaluation shows one side of device suture trap is detached.Functional test was performed on the returned device using foam to replicate the tissue.The needle was able to penetrate thru the foam and pass a stitch.Lever actuation worked as intended.The complaint was not confirmed.There were no indications that would suggest that the device did not meet product specifications upon release into distribution.
|