Model Number 71360046
Device Problem Material Deformation (2976)
Patient Problem No Code Available (3191)
Event Date 10/14/2020
Event Type  Injury  
Event Description
It was reported that during use, outside the patient the screwdriver is stripped.Surgical delay less than 30 minutes.The procedure finished with a competitor device.Patient injuries were not reported.
 
Manufacturer Narrative
Results of investigation: the device, used in treatment, was returned for evaluation.A visual inspection of the returned accord dual ended hex driver confirmed the stated failure.The hex driving tip is rounded and deformed.A hex can strip if the torque applied to the driver exceeds the material strength or if the hex is not seated within the screw when torque is applied.This device exhibits signs of significant use and wear.This device was manufactured in 2017.A medical investigation was conducted and this case reports during use outside of the patient, the screwdriver is stripped.Per complaint description, there was no patient injury, and a delay of less than 30 minutes.The procedure was completed with a competitor device.Since no patient harm is alleged, no further clinical assessment is warranted.A review of complaint history on the listed part revealed no prior complaints for the listed batch with the same failure mode.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ACCORD DUAL ENDED HEX DRIVER
Type of Device
PROSTHESIS, HIP, FEMORAL COMPONENT, CEMENTED, METAL
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10814684
MDR Text Key215461786
Report Number1020279-2020-06314
Device Sequence Number1
Product Code JDG
UDI-Device Identifier03596010611543
UDI-Public03596010611543
Combination Product (y/n)N
PMA/PMN Number
K123598
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Type of Report Initial,Followup
Report Date 11/18/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Model Number71360046
Device Catalogue Number71360046
Device Lot Number17FM05052
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/21/2020
Initial Date Manufacturer Received 10/14/2020
Initial Date FDA Received11/09/2020
Supplement Dates Manufacturer Received11/13/2020
Supplement Dates FDA Received11/18/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-