BD MEDICAL (BD WEST) MEDICAL SURGICAL BD POSIFLUSH NORMAL SALINE SYRINGE; SALINE VASCULAR ACCESS FLUSH Back to Search Results
Catalog Number 306595
Device Problem Delivered as Unsterile Product (1421)
Patient Problem No Patient Involvement (2645)
Event Date 10/16/2020
Event Type  malfunction  
Manufacturer Narrative
A device evaluation is anticipated but has not yet begun.Upon completion of the investigation, a supplemental report will be filed.(b)(4).
 
Event Description
It was reported that the bd posiflush¿ normal saline syringe had no tip cap.The following information was provided by the initial reporter, translated from (b)(6) to english: "when tearing open the package and found that the product has no white cone cap.".
 
Event Description
It was reported that the bd posiflush¿ normal saline syringe had no tip cap.The following information was provided by the initial reporter, translated from chinese to english: "when tearing open the package and found that the product has no white cone cap.".
 
Manufacturer Narrative
The following fields were updated due to additional information: d10: device available for eval yes.D10: returned to manufacturer on: 2020-11-02.H6: investigation summary a device history record review was performed for provided lot number 94101.The review did not reveal any detected quality issues during the production process that could have contributed to this reported incident.To aid in the investigation of this incident, one sample was received for evaluation by our quality team.The sample came with an opened packaging blister and has no tip cap.The sensor at the production line was challenged with the sample received, to which the sensor detected and rejected it.There are sensors at the fill machine, at the assembly and packaging to detect a missing tip cap.A cause for this defect could not be offered at this time.Further action has not been determined necessary at this time.Our quality team will continue to monitor the manufacturing process for this defect and other emerging trends.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BD POSIFLUSH NORMAL SALINE SYRINGE
Type of Device
SALINE VASCULAR ACCESS FLUSH
Manufacturer (Section D)
BD MEDICAL (BD WEST) MEDICAL SURGICAL
1852 10th avenue
columbus NE 68601
MDR Report Key10817306
MDR Text Key230279763
Report Number1911916-2020-01019
Device Sequence Number1
Product Code NGT
Combination Product (y/n)N
PMA/PMN Number
NA
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type foreign,other,user facility
Type of Report Initial,Followup
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Other
Device Expiration Date03/31/2023
Device Catalogue Number306595
Device Lot Number0094101
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/02/2020
Initial Date Manufacturer Received 10/16/2020
Initial Date FDA Received11/10/2020
Supplement Dates Manufacturer Received11/12/2020
Supplement Dates FDA Received11/20/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-