• Decrease font size
  • Return font size to normal
  • Increase font size
U.S. Department of Health and Human Services

MAUDE Adverse Event Report: ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

ABBOTT MEDICAL KIT IMPLANTABLE SLIM TIP LEAD, 50CM; DRG SLIM TIP LEAD Back to Search Results
Model Number MN10450-50A
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problem Cerebrospinal Fluid Leakage (1772)
Event Date 10/08/2020
Event Type  Injury  
Manufacturer Narrative
Date of the event is estimated during processing of this complaint, attempts were made to obtain complete patient information.The results/method and conclusion codes along with investigation results will be provided in the final report.
 
Event Description
It was reported during the procedure, while attempting to place the l5 lead there was a dural puncture.The lead was then placed at l4.The patient experienced a headache which resolved.Therapy was restored.
 
Manufacturer Narrative
Based on the information provided a device problem was not identified, as a result a conclusive cause for the reported patient effect, and the relationship to the product, if any, cannot be determined. .
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
KIT IMPLANTABLE SLIM TIP LEAD, 50CM
Type of Device
DRG SLIM TIP LEAD
Manufacturer (Section D)
ABBOTT MEDICAL
6901 preston rd
plano TX 75024
MDR Report Key10817875
MDR Text Key215558031
Report Number1627487-2020-47625
Device Sequence Number1
Product Code PMP
UDI-Device Identifier05415067027153
UDI-Public05415067027153
Combination Product (y/n)N
PMA/PMN Number
P150004
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative
Type of Report Initial,Followup
Report Date 12/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Expiration Date04/15/2022
Device Model NumberMN10450-50A
Device Catalogue NumberMN10450-50A
Device Lot Number7457529
Was Device Available for Evaluation? No
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/27/2020
Initial Date FDA Received11/10/2020
Supplement Dates Manufacturer Received12/09/2020
Supplement Dates FDA Received12/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-