Catalog Number UNK HIP FEMORAL HEAD
Device Problem Adverse Event Without Identified Device or Use Problem (2993)
Patient Problems Foreign Body Reaction (1868); Hypersensitivity/Allergic reaction (1907); Pain (1994); Test Result (2695); No Code Available (3191)
Event Date 12/03/2018
Event Type  Injury  
Manufacturer Narrative
(b)(4).Initial reporter occupation: lawyer.Follow up is being conducted to determine the legal contact information.If information is obtained that was not available for the initial medwatch, a follow-up medwatch will be filed as appropriate.
 
Event Description
Pinnacle unf and medical records received.After review of medical records the patient was revised to address pain and hypersensitivity.Operative note reported mild metallic staining in the soft tissues and minimal amount of intra-articular fluid.There was a fair amount of back side metallic staining of the liner.Lab result shows metal ions were above 7 ppb.Doi: (b)(6) 2008; dor: (b)(6) 2018: right hip.
 
Manufacturer Narrative
Product complaint # (b)(4).Investigation summary: no device associated with this report was received for examination.The information received will be retained for potential series investigations if triggered by trend analysis, post market surveillance, or other events within the quality system.Depuy considers the investigation closed.Should additional information be received, the information will be reviewed and the investigation will be re-opened as necessary.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
UNK HIP FEMORAL HEAD METAL
Type of Device
METAL FEMORAL HEAD
Manufacturer (Section D)
DEPUY ORTHOPAEDICS INC US
700 orthopaedic drive
warsaw IN 46581 0988
MDR Report Key10818232
MDR Text Key215578638
Report Number1818910-2020-24359
Device Sequence Number1
Product Code KWY
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type consumer,other
Type of Report Initial,Followup
Report Date 10/27/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? No
Device Operator Health Professional
Device Catalogue NumberUNK HIP FEMORAL HEAD
Was Device Available for Evaluation? No
Initial Date Manufacturer Received 10/27/2020
Initial Date FDA Received11/10/2020
Supplement Dates Manufacturer Received11/13/2020
Supplement Dates FDA Received11/16/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Treatment
UNK HIP ACETABULAR LINER METAL PINNACLE; UNK HIP FEMORAL HEAD METAL; UNK HIP FEMORAL STEM SROM; UNK HIP ACETABULAR LINER METAL PINNACLE; UNK HIP FEMORAL HEAD METAL; UNK HIP FEMORAL STEM SROM
Patient Outcome(s) Required Intervention;
Patient Age71 YR
-
-