Device was received and evaluated.Evaluation determined that the device was found with a leak at ¿a-rubber¿ unit.In addition, broken skeleton, metal protruding found in the bending section.The identified parts were replaced, device was repaired.Once completed, the device was tested and passed all required testing and specifications.Root cause of the reported failure likely attributed to user's maintenance and or handling issue.
|
This supplemental report is being submitted to provide the review of the device history records (dhr) and investigation conclusion.Please see updated sections:,g4, g7, h2, h3,h4, h6 and h10.The device history records for this device were reviewed and all records indicated that the product was manufactured according to all applicable procedures and met final product release criteria.No abnormalities were found.A root cause cannot be conclusively determined.However, in instructions for use (ifu),ways of handling which may result in the bending tube¿s breakage and detailed ways of inspections are described.If the user follows the instructions about operation and inspection of the endoscope, the bending tube¿s breakage can be eliminated and any abnormality can be detected safely.The probable cause for this damage is likely that there was deviation from the following description stated in ifu.As stated on the ifu (instruction for use) and as a preventive measure, the user manual states: inspection of the endoscope : inspect the external surface of the entire insertion section, including the bending section and the distal end for dents, bulges, swelling, scratches, peeling of coating, holes, sagging, transformation, bends, adhesion of foreign bodies, missing parts, protruding objects, or other irregularities.In order to prevent the reported event occurrence, ifu ¿instructions for safe use¿ states how to manipulate the device safely.Bending tube breakage can be decreased by manipulating the device in accordance with ifu.Olympus will continue to monitor complaints for this device.
|