LIVANOVA DEUTSCHLAND HEATER-COOLER SYSTEM 3T; CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS Back to Search Results
Model Number 16-02-82
Device Problems Device Sensing Problem (2917); Temperature Problem (3022)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/12/2020
Event Type  malfunction  
Manufacturer Narrative
There was no patient involvement.Livanova (b)(4) manufactures the heater-cooler system 3t.The incident occurred in (b)(6).A review of the dhr could not identify any deviations or nonconformities relevant to the issue.A livanova field service representative was dispatched to the facility to investigate the device and found out also that the device was warming only between 31 and 32 degrees.The temperature sensors were replaced and the heaters placed on order to complete the repair.If any additional information pertinent to the reported event is received, it will be provided in a supplemental report.
 
Event Description
Livanova (b)(4) received a report that a heater-cooler system 3t displayed an error code on the patient side associated to a too high temperature difference of the temperature sensors in the control/safety system during procedure.There was no report of patient injury.
 
Event Description
See initial report.
 
Manufacturer Narrative
H10: the issue was solved by the replacement of the temperature sensor.Subsequent functional verification testing was completed without further issues and the unit was returned to service.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
HEATER-COOLER SYSTEM 3T
Type of Device
CONTROLLER, TEMPERATURE, CARDIOPULMONARY BYPASS
Manufacturer (Section D)
LIVANOVA DEUTSCHLAND
lindberghstr. 25
munich
MDR Report Key10821492
MDR Text Key215885707
Report Number9611109-2020-00625
Device Sequence Number1
Product Code DWC
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional
Remedial Action Other
Type of Report Initial,Followup
Report Date 11/09/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number16-02-82
Was the Report Sent to FDA? No
Initial Date Manufacturer Received 10/12/2020
Initial Date FDA Received11/11/2020
Supplement Dates Manufacturer Received11/10/2020
Supplement Dates FDA Received12/09/2020
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-