MAUDE Adverse Event Report: C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE TEMP SENSING FOLEY CATHETER; TEMPERATURE SENSING FOLEY CATHETER

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

C.R. BARD, INC. (COVINGTON) -1018233 BARDEX LUBRI-SIL ALL-SILICONE TEMP SENSING FOLEY CATHETER; TEMPERATURE SENSING FOLEY CATHETER Back to Search Results
Model Number 119216M
Device Problem Disconnection (1171)
Patient Problem No Consequences Or Impact To Patient (2199)
Event Date 10/16/2020
Event Type  malfunction  
Manufacturer Narrative
The investigation is still in progress.Once the investigation is complete a supplemental report will be filed.
 
Event Description
It was reported that the temperature probe was found lying on the bed out of the foley catheter.
 
Manufacturer Narrative
The reported event was confirmed however the cause was unknown.Visual inspection noted one silicone temperature sensing catheter was received with the meter bag, inlet tubing, and sample port connector.Visual evaluation noted the entire thermistor wire was outside of the lumen and separated from the catheter completely.This fails to meet the specifications as the tip of the thermistor be located according to the corresponding drawing.There was evidence of rtv on the thermistor wire.Although the reported event was confirmed, a root cause could not be determined.A potential root cause for this failure could be insufficient amount of rtv.The device history record was reviewed and found nothing that could have caused or contributed to the reported event.The instructions for use were found adequate and state the following: "recommended inflation capacities 8 fr.And 10 fr.(3cc balloon): use 3.5cc sterile water 12 fr.(5cc balloon): use 5.5cc sterile water 14 fr.And larger (5cc balloon): use 10cc sterile water do not exceed recommended capacities.Warning: after use, this product may be a potential biohazard.Handle and dispose of in accordance with applicable laws and regulations.Caution: aggressive traction, particularly in the presence of suturing, is not recommended for 100% silicone foley catheters to deflate catheter balloon: gently insert a syringe in the catheter valve.Never use more force than is required to make the syringe ¿stick¿ in the valve.If you notice slow or no deflation, re-seat the syringe gently.Allow the balloon to deflate slowly on its own.Do not aspirate or manually accelerate the deflation of the balloon.If permitted by hospital protocol, the valve arm may be severed.If this fails, contact adequately trained professional for assistance, as directed by hospital protocol.Should balloon rupture occur, care should be taken to assure that all balloon fragments have been removed from the patient.Visually inspect the product for any imperfections or surface deterioration prior to use.Note: compatible with appropriate 400-series temperature monitors.Interchangeability + 0.2o c at 37oc.Caution: as with all temperature probes: in the presence of rf energy sources: local heating, temperature errors, and probe damage may occur.In medical use, unplug the temperature-sensing catheter at the extension cable before activating electrosurgical or other types of direct coupled rf energy sources.Do not stretch catheter.This will cause repositioning of probe.Do not use stylet.This will cause stretching of catheter.Corrections: d, h.H11:section a through f - the information provided by bd represents all of the known information at this time.Despite good faith efforts to obtain additional information, the complainant / reporter was unable or unwilling to provide any further patient, product, or procedural details to bd.H3 other text : the actual/suspected device was inspected.
 
Event Description
It was reported that the temperature probe was found lying on the bed out of the foley catheter.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
BARDEX LUBRI-SIL ALL-SILICONE TEMP SENSING FOLEY CATHETER
Type of Device
TEMPERATURE SENSING FOLEY CATHETER
Manufacturer (Section D)
C.R. BARD, INC. (COVINGTON) -1018233
8195 industrial blvd
covington 30014
MDR Report Key10824119
MDR Text Key215950677
Report Number1018233-2020-20890
Device Sequence Number1
Product Code EZL
UDI-Device Identifier00801741046308
UDI-Public(01)00801741046308
Combination Product (y/n)N
PMA/PMN Number
K070582
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type health professional,other,use
Type of Report Initial,Followup
Report Date 12/17/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number119216M
Device Catalogue Number119216M
Device Lot NumberNGET1142
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer10/30/2020
Initial Date Manufacturer Received 10/19/2020
Initial Date FDA Received11/11/2020
Supplement Dates Manufacturer Received12/17/2020
Supplement Dates FDA Received01/07/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Other;
-
-