SMITH & NEPHEW, INC. EVOS SMALL T8 FIXED HANDLE LINEAR DRIVER; SCREWDRIVER Back to Search Results
Model Number 71175077
Device Problem Defective Device (2588)
Patient Problem No Known Impact Or Consequence To Patient (2692)
Event Date 10/21/2020
Event Type  malfunction  
Event Description
It was reported that during a evos small trauma case, the driver has become stripped and is having difficulty capturing screws.Procedure was completed with the same device.No surgical delay or injury to the patient was reported.All available information has been disclosed.If additional information should become available, a supplemental report will be submitted accordingly.
 
Manufacturer Narrative
H3, h6: the associated device, used in treatment, was returned and evaluated.A visual inspection of the returned device confirmed the stated failure mode.The hex end of the device is stripped rendering it inoperable.The device was manufactured in 2018 and shows signs of extensive use.A complaint history review found related failures; this failure mode will be monitored for future complaints and assessed for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.At this time, we do not have reason to suspect that the product failed to meet any product specifications at the time of manufacture.This device is a reusable instrument that can be exposed to numerous surgeries.Damage from prolonged use, misuse or rough handling are likely probable causes of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaint will be reopened.No further investigation is warranted for this complaint; however, we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SMALL T8 FIXED HANDLE LINEAR DRIVER
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10824482
MDR Text Key215885155
Report Number1020279-2020-06378
Device Sequence Number1
Product Code HXX
UDI-Device Identifier00885556636077
UDI-Public00885556636077
Combination Product (y/n)N
PMA/PMN Number
EXEMPT
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 04/29/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? No
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71175077
Device Catalogue Number71175077
Device Lot Number18CUP0006
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Initial Date Manufacturer Received 10/21/2020
Initial Date FDA Received11/11/2020
Supplement Dates Manufacturer Received04/08/2021
Supplement Dates FDA Received04/29/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
-
-