Model Number 71175072
Device Problem Break (1069)
Patient Problem Injury (2348)
Event Date 10/20/2020
Event Type  Injury  
Event Description
It was reported that during surgery (orif ankle), inside the patient, the tip of the hex driver fell off.No delay reported.No patient injuries reported.A s&n backup was available.
 
Manufacturer Narrative
The device, used in treatment, was returned for evaluation.A visual inspection of the returned evos fixed handle confirms the tip of the device is broken off.The broken piece was not returned with the device.This device shows significant signs of wear/ usage.A medical investigation was conducted and this case reports the tip of the hex driver fell off while using it to place/tighten perc guide.Per email communication, the tip remained in perc guide, and nothing was left inside the patient.There was no patient injury or delay, and the procedure was completed using a backup device.Since no patient harm is alleged, no further clinical assessment is warranted.A complaint history review found related failures; this failure mode will be monitored for future complaints for any necessary corrective actions.A review of the manufacturing records did not reveal a manufacturing abnormality that could have caused or contributed to the reported incident.A review of risk management files found that the reported failure was documented appropriately.This device is a reusable instrument that can be exposed to numerous surgeries; damage from repeated use can occur.Expected wear/tear is likely the probable cause of the reported event.We recommend that all reusable instruments be routinely inspected for wear and damage and replaced as necessary.At this time, we have no reason to suspect that the product failed to meet any product specifications at the time of manufacture.Based on this investigation, the need for corrective action is not indicated.Should additional information be received, the complaints will be reopened.No further investigation is warranted for these complaints; however we will continue to monitor for future complaints and investigate as necessary.We consider this investigation closed.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
EVOS SM 2.5MM FIXED HANDLE LINEAR HEX DR
Type of Device
SCREWDRIVER
Manufacturer (Section D)
SMITH & NEPHEW, INC.
1450 brooks rd.
memphis TN 38116
MDR Report Key10826503
MDR Text Key215873666
Report Number1020279-2020-06404
Device Sequence Number1
Product Code HXX
UDI-Device Identifier00885556636046
UDI-Public00885556636046
Combination Product (y/n)N
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,user f
Type of Report Initial,Followup
Report Date 06/24/2021
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Model Number71175072
Device Catalogue Number71175072
Device Lot Number17HUP0001
Was Device Available for Evaluation? Device Returned to Manufacturer
Date Returned to Manufacturer11/11/2020
Initial Date Manufacturer Received 10/20/2020
Initial Date FDA Received11/11/2020
Supplement Dates Manufacturer Received06/23/2021
Supplement Dates FDA Received06/24/2021
Is This a Reprocessed and Reused Single-Use Device? No
Patient Sequence Number1
Patient Outcome(s) Required Intervention;
-
-