Model Number 414008 |
Device Problem
Output Problem (3005)
|
Patient Problem
No Clinical Signs, Symptoms or Conditions (4582)
|
Event Date 10/15/2020 |
Event Type
malfunction
|
Manufacturer Narrative
|
10/19/20: unable to duplicate the reported.Rebooted the entire system, (nexus, generator, gim), verified operation according to service checklist ((b)(4)).The system is fully functional.
|
|
Event Description
|
This incident was reported by a facility in (b)(6) on (b)(6) 2020.The customer states that the fluoro stopped during procedure.The tech had to restart the infimed several times in order to complete the procedure.The reporter also states that the patient was attached to the device during the incident but there was no harm to patient or staff, and that the procedure was completed.
|
|