MAUDE Adverse Event Report: SILK ROAD MEDICAL INC. ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE SDS

  • Print
  • Share
  • E-mail
-
Super Search Devices@FDA
510(k) | DeNovo | Registration & Listing | Adverse Events | Recalls | PMA | HDE | Classification | Standards
CFR Title 21 | Radiation-Emitting Products | X-Ray Assembler | Medsun Reports | CLIA | TPLC
 

SILK ROAD MEDICAL INC. ENROUTE TRANSCAROTID STENT SYSTEM; ENROUTE SDS Back to Search Results
Model Number SR-1030-CS
Device Problems Break (1069); Fracture (1260)
Patient Problem Foreign Body In Patient (2687)
Event Date 10/14/2020
Event Type  Injury  
Manufacturer Narrative
The products associated with this reported event were not returned to the manufacturer for analysis.Nevertheless, the customer provided photographic evidence and a review of the manufacturing records was performed.No issues were identified that could have led to the adverse event reported.Complaints will continue to be reviewed and monitored for trends.A follow-up mdr will be submitted if additional information is obtained.Silk road medical will continue to monitor for occurrences of similar events.On october 23, 2020, a user facility medwatch was received from the hospital directly.A secondary mdr is submitted under 3500a form # 3014526664-2020-00104 since two stents were initially placed on this patient and at this time it is unknown which device may have caused or contributed to the reported event.
 
Event Description
It was reported that after the completion of a transcarotid artery revascularization (tcar) procedure, the patient was complaining of a headache.A computerized tomography scan (ct scan) was performed and revealed the patient experienced filling defect in the ipsilateral cavernous internal carotid artery consistent with a foreign body being present.The patient was transferred to a different hospital where neurosurgery attempted to remove the foreign body via transfemoral approach.On attempts to remove the object, the object inadvertently embolized to the profunda femorous artery, where further attempts to retrieve it were not made.In this position, the foreign body is unlikely to contribute to clinical sequelae.The patient was reported to be well, and without neurological deficit.
 
Search Alerts/Recalls

  New Search  |  Submit an Adverse Event Report

Brand Name
ENROUTE TRANSCAROTID STENT SYSTEM
Type of Device
ENROUTE SDS
Manufacturer (Section D)
SILK ROAD MEDICAL INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer (Section G)
SILK ROAD MEDICAL INC.
1213 innsbruck drive
sunnyvale CA 94089
Manufacturer Contact
magaly boardman
1213 innsbruck drive
sunnyvale, CA 94089
4087309002
MDR Report Key10831669
MDR Text Key216010613
Report Number3014526664-2020-00105
Device Sequence Number1
Product Code NIM
UDI-Device Identifier000811311020492
UDI-Public(01)000811311020492(17)220131(10)17929776
Combination Product (y/n)N
Reporter Country CodeUS
PMA/PMN Number
P140026
Number of Events Reported1
Summary Report (Y/N)N
Report Source Manufacturer
Source Type company representative,health
Reporter Occupation Physician
Type of Report Initial
Report Date 11/12/2020
1 Device was Involved in the Event
1 Patient was Involved in the Event
Is this an Adverse Event Report? Yes
Is this a Product Problem Report? Yes
Device Operator Health Professional
Device Expiration Date01/31/2022
Device Model NumberSR-1030-CS
Device Catalogue NumberSR-1030-CS
Device Lot Number17929776
Was Device Available for Evaluation? No
Is the Reporter a Health Professional? Yes
Initial Date Manufacturer Received 10/15/2020
Initial Date FDA Received11/12/2020
Was Device Evaluated by Manufacturer? Device Not Returned to Manufacturer
Date Device Manufactured01/31/2020
Is the Device Single Use? Yes
Is This a Reprocessed and Reused Single-Use Device? No
Type of Device Usage Initial
Patient Sequence Number1
Patient Outcome(s) Hospitalization; Required Intervention;
Patient Age69 YR
Patient Weight143
-
-