No patient information provided as no patient was involved in this concern.The navlock was returned to the manufacture for evaluation.After functional testing and visual/physical examination the reported issue was confirmed.The returned tracker has areas of galling inside the handle not allowing the insertion of known good instruments.Otherwise, with markers attached and fully seated, the tracker displays a good geometry error with normal tracking.If information is provided in the future, a supplemental report will be issued.
|